Ultrasonic Flow Meter

Description

  • Hand-held Ultrasonic Type

It works on the transit time method. This is based on the principle that sound waves traveling with the flow will move faster than those traveling against it. The resulting difference in transit time is directly proportional to the flow velocity of the liquid and consequently to the flow rate.

  • Portable Ultrasonic Type

It is available in a variety of configuration that permit the user to select an ultrasonic meter with feature suitable to meet particular application requirements. It could also provides the data printed service. Built-in min thermal printed with instant and timing print function and uplink over 20 measuring data to computer or internet.

  • Wall mounted Ultrasonic Type

It is a fixed mounted transit-time ultrasonic flowmeter, with clamp-on transducers for non-invasive liquid measurement. Our microprocessor based, user friendly, field programmable flow measurement technique allows no interruption of the process flow and has low installation cost.

Description

Features

  • The ultrasonic flow meters are used in a closed conduit. The transducers are a non-contacting, clamp-on type, which will provide benefits of non-fouling operation and easy installation.
  • They can be virtually applied to a widely range of measurements. The measured pipe ranges 15-6000mm. Because the instrument and transducers are non-contacting and no moving parts.
  • The series transit-time flow meter utilizes two transducers that function as both ultrasonic transmitters and receivers. The transducers are clamped on the outside of a closed pipe at a specific distance from each other. The transducers can be mounted in V-method where the sound transverses the pipe twice, or W-method where the sound transverses the pipe four times, or in Z-method where the transducers are mounted on opposite sides of the pipe and the sound crosses the pipe once. This selection of the mounting method depends on pipe and liquid characteristics. The flow meter operates by alternately transmitting and receiving a frequency modulated burst of sound energy between the two transducers and measuring the transit time that it takes for sound to travel between the two transducers.

Specifications

Liquid TypesMost clean liquids or liquids containing small amounts of suspended solids or gas bubbles
Measuring PrincipleTransit-Time
Converter ModelTUF-2000PPortable with Printer
TUF-2000HHand-Held
TUF-2000SWall-Mounted
Pipe SizeDN15...DN6000
Sensor ModelTS-2DN15...DN100
TM-1DN50...DN700
TL-1DN300...DN6000
HTS-1DN15...DN100
HTM-1DN50...DN700
Max. Fluid TemperatureTS-2;TM-1;TL-1:-40...+90℃
HTS-1,;HTM-1:-40...+160℃
Accuracy±1%~±2% value of reading (0.5-30m/s)
±0.5% value of reading(online    calibration)
Power Supply and Output(Depending on Model)(1) Rechargeable Battery(RS232)
(2) 110-230Vac(4-20mA/Pulse/RS485)
(3) 24V DC(4-20mA/Pulse/RS485)
Pipe MaterialsCast Iron; Stainless Steel
Ductile Iron Copper; PVC; Aluminum
Asbestos Fiberglass...etc
Liner MaterialsTar Epoxy, Rubber, Morta
Polypropylene, Polystyrol
Polystyrene, Polyester, Ebonite
Polyethylene, Teflon...etc
LanguageEnglish; Chinese(Other's on request)
Engineer UnitsM3;Liter;US Gallon; Imperial
Gallon; Million Gallon; Cubic Feet
US Barrels; Imperial Barrels; Oil    Barrel
Totalizer7 digit; Forward; Reverse & Net    Values
Flow Rate5 digit with decimal point
Host MaterialCast Aluminium
WeightAround 7 KG/PCS